C-Pace EP Culture Pacer
The C-Pace EP is a multi-channel stimulator designed for the chronic stimulation of bulk quantities of cells in culture incubators. This pacer will emit bipolar stimuli, greatly reducing problems with electrolysis at the electrodes. It has an output voltage of +/- 40 V, a frequency from once every 100 seconds to 99 pulses/sec, and duration 0.4 to 24 msec. The EP has a flexibility in its pacing protocols that has been designed on request by customers who want to study the effects of arrhythmia or exercise protocols during culture. Gate and Aux outputs allow for increased functionality when used as an acute pacer.
Extra Features in the EP version:
- Arrhythmia Protocols.
- Insert an offbeat pulse at a fixed interval.
- Exercise Protocols.
- Run multiple pulse trains with individually programmable frequency and duration.
- Irregular Pacing.
- Random variation of specified frequency within definable percentage window and guaranteed average effective rate.
- External Triggering.
- Change to next pulse train with TTL ‘advance’ input and/or trigger individual pulses with TTL ‘pulse’ input.
| Type | Title | Size | Date |
PDF | C-Pace EP Hardware Manual | 300 kbs | 11th January 2010 |